Objective: To perform an annual test on the system stability of the first helical tomotherapy unit (Tomotherapy, TOMO) in Zhejiang province according to WS531-2017 "specification for testing of quality control in helical tomotherapy unit", and to further standardize the quality control operation, understand the properties of the system and ensure the accurate implementation of clinical radiotherapy. Methods: According to the requirements of WS531-2017, 8-channel measurement instrument (TomoElectrometer), ionization chamber A1SL (Standing Imaging, USA) and A17 (Standing Imaging, USA), cylindrical phantom (Cheese Phantom), equivalent rectangular solid water, two-dimensional water tank, EBT3 film, and Vidar film analyzer were used to detect 10 key indicators of TOMO. Results: The static output dose deviation was -0.6%, and the rotational output dose deviation was 1.4%. The difference between radiation quality and planned value was 0.8%. The symmetry of the horizontal dose curve of the radiation field was -1.2%; and the full width at half-maximum deviation of the longitudinal dose curve of the radiation field was 0.7mm. The horizontal offset of multi-leaf collimator was -0.6mm. The green laser offset in axial plane, sagittal plane and coronal plane were 0.08mm, 0.2mm and 0.2mm to the of virtual isocenter. The red laser light offset was -0.4mm. The movement deviation of the treatment bed was 0.3mm; and the synchronous deviation of the rotation of the treatment bed and frame was -0.6mm. Conclusion: All 10 key indicators of TOMO meet the requirement of WS531-2017. Some necessary testing items, such as MVCT image quality verification, should be added to the annual TOMO quality control in order to ensure the good stability and normal working of system.
Published in | International Journal of Clinical and Experimental Medical Sciences (Volume 5, Issue 1) |
DOI | 10.11648/j.ijcems.20190501.14 |
Page(s) | 14-18 |
Creative Commons |
This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited. |
Copyright |
Copyright © The Author(s), 2019. Published by Science Publishing Group |
Helical Tomotherapy Unit, Quality Assurance, Radiotherapy Dose, Deviation
[1] | Li Y, Xu H J. The modern physics of radiation oncology [M]. Beijing: China Atomic Energy Press, 2015. |
[2] | Sen A, West M K. Commissioning experience and quality assurance of helical tomotherapy machines [J]. Journal of Medical Physics, 2009, 34(4): 194-199. |
[3] | Mackie T R. History of tomotherapy [J]. Physics in Medicine & Biology, 2006, 51(13): 427-453. |
[4] | Ruchala K J, Olivera G H, Schloesser E A, et al. Megavoltage CT on a tomotherapy system. [J]. Physics in Medicine & Biology, 1999, 44(10): 2597-2603. |
[5] | Ricchetti F, Barra S, Agostinelli S, et al. Feasibility of helical tomotherapy for radical dose retreatment in pelvic area: a report of 4 cases.[J]. Tumori, 2011, 97(4): 492. |
[6] | Chen Q, Westerly D, Fang Z, et al. TomoTherapy MLC verification using exit detector data [J]. Medical Physics, 2012, 39(1): 143-151. |
[7] | Langen K M, Papanikolaou N, Balog J, et al. QA for helical tomotherapy: Report of the AAPM Task Group 148a [J]. Medical Physics, 2010, 37(9): 4817-4853. |
[8] | Jeraj R, Mackie T R, Balog J, et al. Dose calibration of nonconventional treatment systems applied to helical tomotherapy [J]. Medical Physics, 2005, 32(2): 570-577. |
[9] | Kapatoes J M, Olivera G H, Ruchala K J, et al. A feasible method for clinical delivery verification and dose reconstruction in tomotherapy [J]. Medical Physics, 2001, 28(4): 528-542. |
[10] | Chen Q, Levinson L, Ding K, et al. TomoTherapy Patient QA Using Exit Detector Measurement of Pre-Treatment In-Air Delivery [J]. Medical Physics, 2012, 39(9): 3707-3707. |
[11] | Chen Q, Westerly D, Fang Z, et al. TomoTherapy MLC verification using exit detector data. [J]. Medical Physics, 2012, 39(1): 143-151 |
[12] | Fenwick J D, Tomé W A, Jaradat H A, et al. Quality assurance of a helical tomotherapy machine [J]. Physics in Medicine & Biology, 2004, 49 (13): 2933-2953. |
[13] | Klein E E, Joseph H, John B, et al. Task Group 142 report: Quality assurance of medical accelerators [J]. Medical Physics, 2009, 36(9): 4197-4207. |
[14] | Fenwick J D, Tomé W A, Jaradat H A, et al. Quality assurance of a helical tomotherapy machine. [J]. Physics in Medicine & Biology, 2004, 49(13): 2933-2953. |
[15] | Jeraj R, Mackie T R, Balog J, et al. Radiation characteristics of helical tomotherapy [J]. Medical Physics, 2004, 31(2): 396-403. |
APA Style
Cheng Xiaolong, Liu Jiping, Shi Jianfang. (2019). Implementation and Evaluation of WS531-2017 "Specification for Testing of Quality Control in Helical Tomotherapy Unit". International Journal of Clinical and Experimental Medical Sciences, 5(1), 14-18. https://doi.org/10.11648/j.ijcems.20190501.14
ACS Style
Cheng Xiaolong; Liu Jiping; Shi Jianfang. Implementation and Evaluation of WS531-2017 "Specification for Testing of Quality Control in Helical Tomotherapy Unit". Int. J. Clin. Exp. Med. Sci. 2019, 5(1), 14-18. doi: 10.11648/j.ijcems.20190501.14
AMA Style
Cheng Xiaolong, Liu Jiping, Shi Jianfang. Implementation and Evaluation of WS531-2017 "Specification for Testing of Quality Control in Helical Tomotherapy Unit". Int J Clin Exp Med Sci. 2019;5(1):14-18. doi: 10.11648/j.ijcems.20190501.14
@article{10.11648/j.ijcems.20190501.14, author = {Cheng Xiaolong and Liu Jiping and Shi Jianfang}, title = {Implementation and Evaluation of WS531-2017 "Specification for Testing of Quality Control in Helical Tomotherapy Unit"}, journal = {International Journal of Clinical and Experimental Medical Sciences}, volume = {5}, number = {1}, pages = {14-18}, doi = {10.11648/j.ijcems.20190501.14}, url = {https://doi.org/10.11648/j.ijcems.20190501.14}, eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.ijcems.20190501.14}, abstract = {Objective: To perform an annual test on the system stability of the first helical tomotherapy unit (Tomotherapy, TOMO) in Zhejiang province according to WS531-2017 "specification for testing of quality control in helical tomotherapy unit", and to further standardize the quality control operation, understand the properties of the system and ensure the accurate implementation of clinical radiotherapy. Methods: According to the requirements of WS531-2017, 8-channel measurement instrument (TomoElectrometer), ionization chamber A1SL (Standing Imaging, USA) and A17 (Standing Imaging, USA), cylindrical phantom (Cheese Phantom), equivalent rectangular solid water, two-dimensional water tank, EBT3 film, and Vidar film analyzer were used to detect 10 key indicators of TOMO. Results: The static output dose deviation was -0.6%, and the rotational output dose deviation was 1.4%. The difference between radiation quality and planned value was 0.8%. The symmetry of the horizontal dose curve of the radiation field was -1.2%; and the full width at half-maximum deviation of the longitudinal dose curve of the radiation field was 0.7mm. The horizontal offset of multi-leaf collimator was -0.6mm. The green laser offset in axial plane, sagittal plane and coronal plane were 0.08mm, 0.2mm and 0.2mm to the of virtual isocenter. The red laser light offset was -0.4mm. The movement deviation of the treatment bed was 0.3mm; and the synchronous deviation of the rotation of the treatment bed and frame was -0.6mm. Conclusion: All 10 key indicators of TOMO meet the requirement of WS531-2017. Some necessary testing items, such as MVCT image quality verification, should be added to the annual TOMO quality control in order to ensure the good stability and normal working of system.}, year = {2019} }
TY - JOUR T1 - Implementation and Evaluation of WS531-2017 "Specification for Testing of Quality Control in Helical Tomotherapy Unit" AU - Cheng Xiaolong AU - Liu Jiping AU - Shi Jianfang Y1 - 2019/05/15 PY - 2019 N1 - https://doi.org/10.11648/j.ijcems.20190501.14 DO - 10.11648/j.ijcems.20190501.14 T2 - International Journal of Clinical and Experimental Medical Sciences JF - International Journal of Clinical and Experimental Medical Sciences JO - International Journal of Clinical and Experimental Medical Sciences SP - 14 EP - 18 PB - Science Publishing Group SN - 2469-8032 UR - https://doi.org/10.11648/j.ijcems.20190501.14 AB - Objective: To perform an annual test on the system stability of the first helical tomotherapy unit (Tomotherapy, TOMO) in Zhejiang province according to WS531-2017 "specification for testing of quality control in helical tomotherapy unit", and to further standardize the quality control operation, understand the properties of the system and ensure the accurate implementation of clinical radiotherapy. Methods: According to the requirements of WS531-2017, 8-channel measurement instrument (TomoElectrometer), ionization chamber A1SL (Standing Imaging, USA) and A17 (Standing Imaging, USA), cylindrical phantom (Cheese Phantom), equivalent rectangular solid water, two-dimensional water tank, EBT3 film, and Vidar film analyzer were used to detect 10 key indicators of TOMO. Results: The static output dose deviation was -0.6%, and the rotational output dose deviation was 1.4%. The difference between radiation quality and planned value was 0.8%. The symmetry of the horizontal dose curve of the radiation field was -1.2%; and the full width at half-maximum deviation of the longitudinal dose curve of the radiation field was 0.7mm. The horizontal offset of multi-leaf collimator was -0.6mm. The green laser offset in axial plane, sagittal plane and coronal plane were 0.08mm, 0.2mm and 0.2mm to the of virtual isocenter. The red laser light offset was -0.4mm. The movement deviation of the treatment bed was 0.3mm; and the synchronous deviation of the rotation of the treatment bed and frame was -0.6mm. Conclusion: All 10 key indicators of TOMO meet the requirement of WS531-2017. Some necessary testing items, such as MVCT image quality verification, should be added to the annual TOMO quality control in order to ensure the good stability and normal working of system. VL - 5 IS - 1 ER -